Helium-Neon Laser Associated with Laserscope.
Abstract:
An optical radiation special study of the Laserscope was performed on the Laserscope during August 1981. It was determined that the 2.5 mW unit posed a hazard to the unprotected eye within 10 m even for momentary viewing 0.25 s. The 0.5 mW units did not present this hazard, assuming the output does not exceed 1 mW no 0.5 mW units were available for measurement. Intrabeam viewing of the 2.5 mW unit should not be permitted within 135 m for lengthy time periods. Intrabeam viewing with optical instruments should not be permitted within 150 m. Author