Calibration History of Some Rotronic MP-1OO and Vaisala Humicap Relative Humidity Sensors.

reportActive / Technical Report | Accession Number: ADA296674 | Open PDF

Abstract:

An analysis has been made of the calibrations done during 1990 through 1994 on Rotronic MP-1OO relative humidity sensors used in the Improved METeorological IMET system and sensors built at the Woods Hole Oceanographic Institution WHOl with the Vaisala Humicap sensing element used in the Vector Averaging Wind Recorder VAWR. The shift from one calibration to the next is, typically, 2-3 RH which represents the major uncertainty in relative humidity with either of these sensors. The direction of each shift appears to be random thus, there does not appear to be any long-term drifts.