Time-Dependent and Light-Induced Fading in Victoreen Model 2600-80 Aluminum Oxide Thermoluminescence Dosemeters
Abstract:
In the absence of light, the thermoluminescence TL signal of irradiated aluminum oxide TL dosemeters TLD faded by 21 in 3 months. When exposed to low intensity incandescent light. the TL signal decreased by 25 in 6 h. High intensity fluorescent light caused the TL signal to fade by more than 90 in 6 h.