Test and Evaluation Report of the IMED Volumetric Infusion Pump Model 960A

reportActive / Technical Report | Accession Number: ADA248538 | Open PDF

Abstract:

The IMED Volumetric Infusion Pump Model 960A was tested for electromagnetic interferencecompatibility in the UH-60A helicopter under the U. S Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interferencecompatibility and human factors. The IMED Volumetric Infusion Pump Model 960A was found to be compatible with the U.S. Army medical evacuation UH-60A Blackhawk.