Dose Equivalent LiF TLD (Thermoluminescent Dosimeter) Neutron and Photon Area Monitor

reportActive / Technical Report | Accession Number: ADA211901 | Open PDF

Abstract:

A passive area monitor has been demonstrated that can measure neutron and gamma-ray dose equivalents. The Monitor combines the neutron moderator shield of a Snoopy remeter with a Lithium Fluoride TLD thermal neutron and gamma-ray detector. This area monitor can be a valuable aid in evaluating personnel LiF TLD albedo neutron dosimeter exposures. Calibrated with PuBe, the Monitor will read the actual neutron dose equivalent -60 160 percent over the neutron energy range from thermal to 14 MeV. Calibrated with Cobalt 60, the Monitor will read the actual gamma-ray dose equivalent from -40 to 0 percent from 38 keV to Cobalt 60. Keywords Radiation monitor, Photons, Remeter.