Hadamard Transform Spectrometer

reportActive / Technical Report | Accession Number: AD0740585 | Open PDF

Abstract:

A Hadamard Transform Spectrometer is described. This multiplexing spectrometer is based on a 127 square cyclic Hadamard matrix. A resolving power of 0.1cm is achieved with a multiplexing gain factor of over five-fold in signal-to-noise ratio over use of a conventional single exit slit. A procedure for calibrating an infrared prism spectrometer plots drum numbers and wavelengths as ordinates against a trigonometric function of the index of the prism, as parameter and abscissae. The calibration data yields a straight line so only calibration data for two wavelengths is necessary.

Security Markings

DOCUMENT & CONTEXTUAL SUMMARY

Distribution:
Approved For Public Release, Document Partially Illegible
Distribution Statement:
Approved For Public Release; Distribution Is Unlimited. Document Partially Illegible.

RECORD

Collection: TR
Identifying Numbers
Subject Terms