Instrumented Impact Tester Operation Procedure for Izod Specimen.

reportActive / Technical Report | Accession Number: AD0728805 | Need Help?

Abstract:

A procedure was developed using instrumented Izod cantilever beam Impact Tester. This procedure can be used for testing various types of material such as plastics, light metals, ceramics, and materials used in the non-metallic cartridge cases. There are two hammers available at present. A two foot pound, and 10 foot pound. The height of release of the pendulum is two feet, giving a striking velocity of approximately eleven foot pound seconds. An oscilloscope with storage display is used to record the results. Author

Security Markings

DOCUMENT & CONTEXTUAL SUMMARY

Distribution:
Approved For Public Release

RECORD

Collection: TR
Identifying Numbers
Subject Terms