DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
AD0883290
Title:
Expansion Technique for Cary 14 Transittance Measurements.
Corporate Author:
NAVAL WEAPONS CENTER CHINA LAKE CA
Report Date:
1971-04-01
Abstract:
The expansion feature of the transmittance T slidewire on the Cary Model 14 spectrophotometer, 0 to 1 T Full Scale, provides an extremely useful technique to measure the magnitude of small, yet significant energy leaks 1 T of sidesbands in interference filters. Of the several optical materials investigated, the optical wedge provided to be the best component to establish the 0 to 1 T range. It is insensitive to temperature changes, has excellent shelf life, is commercially available at a reasonable cost, and can be used over a wide wavelength range. In a test with a neutral density filter, good agreement existed between the values from the 0 to 1 T range and from the 0 to 100 T and 0 to 10 T ranges. The expansion technique takes advantage of the excellent photometric accuracy of the spectrophotometer by providing an improved readability of chart values. Author
Descriptive Note:
Technical publication,
Pages:
0013
File Size:
0.00MB