Abstract:
To determine impurity concentrations in solids using mass spectographs, ion-sensitive emulsions are used. The principal problem in obtaining quantitative results from a mass spectrum recorded on a photographic plate is determining the characteristic response curve. Calculations required to determine the parameters in an empirical function which accurately represents the response curve of Ilford Q2 emulsions are programmed in FORTRAN language. The program input data consists of the impurity-ion mass, line density, and exposure the impurity-ion concentration and the detection limit are calculated and printed as output.