Accession Number:

ADA614660

Title:

Prehospital Emergency Inguinal Clamp Controls Hemorrhage in Cadaver Model

Descriptive Note:

Journal article

Corporate Author:

ARMY INST OF SURGICAL RESEARCH FORT SAM HOUSTON TX

Report Date:

2013-07-01

Pagination or Media Count:

8.0

Abstract:

Background The Combat Ready Clamp is indicated to stop difficult inguinal bleeding on the battlefield, the most common type of junctional bleeding and now the most common cause of preventable battlefield death. The purpose of the present study is to report the data of clamp development to help appliers use it correctly. Methods Wake Forest University investigators used a cadaver model to test the clamp s ability to control hemorrhage. Ten fresh cadavers were made to simulate inguinal and popliteal wound bleeding. Blood simulant was pumped to quantify device effectiveness in testing. Points of application included proximal pressure point control of popliteal, inguinal, and bilateral bleeding. Results Clamp use promptly controlled pulsing arterial hemorrhages from inguinal, popliteal, and bilateral wounds. The device, when placed on the common iliac artery, stopped all ipsilateral distal bleeding. Conclusions The evidence of how the clamp works in the cadaver model showed that clamp use can plausibly be tailored to control inguinal hemorrhage from one wound, control two ipsilateral wounds with hemorrhage from one artery e.g., common iliac artery, and control bilateral inguinal wounds compression of the origins of bilateral common iliac arteries.

Subject Categories:

  • Medicine and Medical Research
  • Medical Facilities, Equipment and Supplies

Distribution Statement:

APPROVED FOR PUBLIC RELEASE