DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
ADA604286
Title:
Gamma Imaging using Rotational Modulation Collimation
Descriptive Note:
Technical rept.
Corporate Author:
DEFENCE SCIENCE AND TECHNOLOGY ORGANISATION VICTORIA (AUSTRALIA)
Report Date:
2014-01-01
Pagination or Media Count:
27.0
Abstract:
This report describes measurements undertaken at the DSTO radiation laboratory to characterise the performance of two rotating modulation collimator RMC gamma imagers built by DSTO. The ability of these devices to image shielded and unshielded Co-60 and Cs-137 sources is demonstrated. A discussion of the characteristics of this and other gamma imaging techniques and their application in a Defence context is included.
Distribution Statement:
APPROVED FOR PUBLIC RELEASE