Accession Number:

ADA411224

Title:

Behavior of an Automatic Pacemaker Sensing Algorithm for Single-Pass VDD Atrial Electrograms

Descriptive Note:

Corporate Author:

MYONGII UNIV YONGIN (KOREA) DEPT OF ELECTRONICS ENGINEERING

Report Date:

2001-10-25

Pagination or Media Count:

4.0

Abstract:

Single-pass VDD pacemakers have been used as a result of simple implantation procedures and generally reliable atrial tracking. However, there is a controversy over their reliabilities of atrial tracking. As a new sensing method for reliable atrial tracking, a simple automatic pacemaker sensing algorithm was implemented and evaluated to validate its benefits in sensing depolarization waves of singlepass VDD atrial electrograms. The automatic sensing algorithm had a predetermined sensing dynamic range and the sensitivity level was controlled as 50 of the average of two most recently sensed intrinsic amplitudes. The behavior of the automatic sensing algorithm in the single-pass VDD atrial electrograms was analyzed and characterized. It was observed that the automatic sensing algorithm was more effective than a conventional fixed threshold method to accurately detect and track p-waves in SVDD electrograms.

Subject Categories:

  • Medicine and Medical Research
  • Medical Facilities, Equipment and Supplies

Distribution Statement:

APPROVED FOR PUBLIC RELEASE