Abstract:

The Physio Control DefibrillatorMonitor Model LIFEPAK 10 was tested for electromagnetic interferencecompatibility in the UH-60A helicopter under the U.S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interferencecompatibility and human factors. The LIFEPAK 10 passed the overall evaluation and is validated as compatible with U.S. Army aeromedical aircraft.