DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
ADA193192
Title:
Radiometry Using Thermal Images. Part 2. Technical Details.
Descriptive Note:
Technical rept.,
Corporate Author:
ELECTRONICS RESEARCH LAB ADELAIDE (AUSTRALIA)
Report Date:
1987-09-01
Pagination or Media Count:
25.0
Abstract:
In the first part of this report the general operating procedures and system description of the AGA Model 782 Thermovision were outlined. The calibration of the system to determine the relationship between the Thermal value an empirical quantity related to the energy received and temperature was described in detail. To enable the radiometric use of these thermal imagers, characteristics such as Field of View, Spectral Response and Irradiance Sensitivity had to be measured in order to validate the manufacturerss supplied specifications. The increased emphasis on the need to enhance the digitized images using image processing techniques has demanded the precise measurement of both the horizontal and vertical Modulation Transfer Functions in order to ascertain optimum sampling rates in both dimensions. This report will describe the techniques used to measure each of the necessary performance characteristics, detail all performance specifications of the Short Waveband SWB and Long Waveband LWB Model 782 Thermovisions in use in IOC Group in all hardware configurations and describe the optical path, scanning principle and video format produced by the systems.
Distribution Statement:
APPROVED FOR PUBLIC RELEASE