DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
ADA188686
Title:
Flywheel-Powered Mobile X-Ray Generator.
Descriptive Note:
Final rept. 1 Jan 82-31 Oct 85,
Corporate Author:
WISCONSIN UNIV-MADISON
Report Date:
1985-10-15
Pagination or Media Count:
110.0
Abstract:
Medical x ray machines require high power for good images without motion artifacts. Power requirements are 5 kWp for small image fields and 15 kWp for larger fields, beyond the capacity of a small field generator. Conventional batteries and high voltage capacitors can be used to accumulate energy but have practical limitations of size or affect x-ray beam quality. Circuits and mechanical components have been developed which use small flywheels with certain alternators. Aircraft alternators were used as both motors and generators to accumulate energy at a slow rate and release the energy in a short burst suitable for operating a diagnostic x-ray machine. The flywheel alternator assembly accumulated energy from a 0.5 kW source and released it in bursts in excess of 35 kWp regulated and 50 kWp unregulated. Keywords X ray, Energy storage, Flywheel powered.
Distribution Statement:
APPROVED FOR PUBLIC RELEASE