Accession Number:

ADA188686

Title:

Flywheel-Powered Mobile X-Ray Generator.

Descriptive Note:

Final rept. 1 Jan 82-31 Oct 85,

Corporate Author:

WISCONSIN UNIV-MADISON

Personal Author(s):

Report Date:

1985-10-15

Pagination or Media Count:

110.0

Abstract:

Medical x ray machines require high power for good images without motion artifacts. Power requirements are 5 kWp for small image fields and 15 kWp for larger fields, beyond the capacity of a small field generator. Conventional batteries and high voltage capacitors can be used to accumulate energy but have practical limitations of size or affect x-ray beam quality. Circuits and mechanical components have been developed which use small flywheels with certain alternators. Aircraft alternators were used as both motors and generators to accumulate energy at a slow rate and release the energy in a short burst suitable for operating a diagnostic x-ray machine. The flywheel alternator assembly accumulated energy from a 0.5 kW source and released it in bursts in excess of 35 kWp regulated and 50 kWp unregulated. Keywords X ray, Energy storage, Flywheel powered.

Subject Categories:

  • Electric Power Production and Distribution
  • Medical Facilities, Equipment and Supplies

Distribution Statement:

APPROVED FOR PUBLIC RELEASE