Title:

Response Characteristics of the Neil Brown Instrument Systems, Inc. mark III CTD to Step Changes in Temperature and Conductivity.

Descriptive Note:

Final rept.,

Corporate Author:

NAVAL OCEANOGRAPHIC OFFICE NSTL STATION MS

Personal Author(s):

Report Date:

1980-10-01

Pagination or Media Count:

18.0

Abstract:

Response characteristics of the Mark 3 CTD temperature and conductivity sensors have been determined. A 45 degree rotation of the thermistor bead assembly into the path of flow provides maximum dynamic response. The time constant for the temperature sensor was measured to be approximately 45 milliseconds while that of the conductivity sensor was measured to be approximately 20 milliseconds.