DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
ADA090105
Title:
Development and Characterization of a 9-mm Inductively Coupled Argon Plasma (ICP) Source for Atomic Emission Spectrometry.
Descriptive Note:
Interim technical rept.,
Corporate Author:
INDIANA UNIV AT BLOOMINGTON DEPT OF CHEMISTRY
Report Date:
1980-09-30
Pagination or Media Count:
34.0
Abstract:
A new 9-mm i.d. inductively coupled plasma torch is described which supports a stable, analytically useful plasma at less than 500 watts of r.f. power and 7 Lmin total argon gas flow. Detection limits, working curves and other analytical characteristics of the new device are compared with those of both a miniature 13-mm i.d. and conventional 19-mm i.d. ICP. Although temperatures of the new plasma are somewhat lower than those in the larger plasmas, the new system offers promise for future, miniaturized ICP instruments. Author
Distribution Statement:
APPROVED FOR PUBLIC RELEASE