Accession Number:

ADA065094

Title:

Elemental Analysis of Gas Chromatographic Effluents With An Inductively Coupled Plasma.

Descriptive Note:

Technical rept.,

Corporate Author:

ARIZONA UNIV TUCSON DEPT OF CHEMISTRY

Personal Author(s):

Report Date:

1979-02-01

Pagination or Media Count:

25.0

Abstract:

The performance of the inductively-coupled plasma ICP as an element selective gas chromatographic GC detector is investigated. Special emphasis is placed on evaluating the ability of the ICP to perform simultaneous multi-element analyses on each component of a mixture. In general, ICP detection limits, linear dynamic ranges and selectivities are shown to compare favorably with those which have been observed for the flame photometric detectors FPD and microwave-excited plasma detectors MEPD. Author

Subject Categories:

  • Atomic and Molecular Physics and Spectroscopy

Distribution Statement:

APPROVED FOR PUBLIC RELEASE