Abstract:

Linearity measurements and integral pulse height distributions have been used to evaluate the effects of electron multiplier voltage and discriminator coefficient Ad on the stability, sensitivity and dynamic range of three pulse counting PC measurement systems. The data show that selection of the highest possible voltage has significant advantages for most experiments, but that the choice of an optimum value for Ad depends on the requirements of the particular experiment and requires careful evaluation of the trade-offs among stability, sensitivity, and dynamic range. Semiquantitative evaluation of these trade-offs on the basis of the experimental results is illustrated. The data also show clearly that measurements such as those reported here must be made on each system, since specific results for one system can not be generalized to others. Author

Subject Categories:

  • Electrical and Electronic Equipment
  • Test Facilities, Equipment and Methods

Distribution Statement: