Accession Number:

ADA043565

Title:

Improving the Analytical Accuracy of Electron Spin Resonance Spectrometry.

Descriptive Note:

Technical rept. no. 10, 1 Jul-20 Aug 77,

Corporate Author:

ROCKWELL INTERNATIONAL THOUSAND OAKS CALIF SCIENCE CENTER

Personal Author(s):

Report Date:

1977-08-22

Pagination or Media Count:

15.0

Abstract:

Sources of error in Analytical Electron Spin Resonance Spectrometry are discussed. Errors due to gain of the lock in amplifiers and modulation amplitude settings can be reduced if the settings are properly calibrated or if the attenuators are rebuilt. It is also necessary that the incident microwave power level is monitored and that the cavity is reproducibly tuned. A method of correction for incomplete integration of Lorentzian lines is discussed.

Subject Categories:

  • Electrical and Electronic Equipment
  • Electricity and Magnetism

Distribution Statement:

APPROVED FOR PUBLIC RELEASE