Accession Number:

ADA038271

Title:

Production Engineering Measure for High Range Geiger Mueller Tube

Descriptive Note:

Final rept. May 1974-Sep 1976

Corporate Author:

LND INC OCEANSIDE NY

Personal Author(s):

Report Date:

1976-09-01

Pagination or Media Count:

109.0

Abstract:

This program is an investigation of the proper techniques employed to provide stable, predictable performance GM tubes for use as the high range detector of the ANVDR-1. The techniques include special cathode and anode surface treatment to prevent secondary emissions and resultant spurious counts instability. All testing has been completed during this reporting period. The tests prove the units to be completely stable and meet all the requirements of the specification.

Subject Categories:

  • Nuclear Instrumentation

Distribution Statement:

APPROVED FOR PUBLIC RELEASE