Title:

Spectral Sensitivity Calibration of Spectrophotometers,

Descriptive Note:

Corporate Author:

ADMIRALTY MATERIALS LAB POOLE (ENGLAND)

Personal Author(s):

Report Date:

1976-03-01

Pagination or Media Count:

11.0

Abstract:

A standard lamp calibrated by NPL was used to measure the spectral sensitivity curve of a spectrophotometer which had previously been calibrated in the ultraviolet region against the ferrioxalate actinometer. The results indicated that the calibration of the lamp can be extrapolated appreciably below 300 nm with sufficient accuracy for many purposes. The spectral sensitivity curve of a previously uncalibrated double grating spectrophotometer was measured by this technique. Author

Subject Categories:

  • Test Facilities, Equipment and Methods
  • Optics