Accession Number:

ADA024224

Title:

Calorimetric Method of Sensitivity Calibration for Far-Infrared Detectors.

Descriptive Note:

Interim rept.,

Corporate Author:

NAVAL RESEARCH LAB WASHINGTON D C

Personal Author(s):

Report Date:

1976-04-16

Pagination or Media Count:

19.0

Abstract:

A calorimetric method for measuring the absolute spectral sensitivity of a far-infrared 100 micrometers to 5 mm detector was developed. The method uses a liquid-helium-cooled integrating-sphere calorimeter capable of measuring a radiative power of the order of 1 microwatt. A preliminary sensitivity measurement of an InSb detector at 3.4 mm has been made. Author

Subject Categories:

  • Physical Chemistry
  • Infrared Detection and Detectors

Distribution Statement: