Transient Calorimeter Calibration System

Descriptive Note:

Final rept. 1 Mar 1973-1 Feb 1974

Corporate Author:

TAMARACK SCIENTIFIC CO ORANGE CA

Personal Author(s):

Report Date:

1975-03-01

Pagination or Media Count:

56.0

Abstract:

Null point calorimeters are used to measure heat flux levels in excess of 15.0 kWper sq cm in arc heated re-entry wind tunnels. To augment these measurements at the Air Force Flight Dynamics Laboratory a calibration system and a transfer standard was developed. The system generates heat flux densities up to 10 kWper sq cm within a 0.25 in. square target zone using a xenon arc radiation source. The absorptivity of calorimeter surfaces to this radiant source is measured with an accessory apparatus to yield a calibration accuracy of 5.