Accession Number:

AD1022078

Title:

Feasibility of a Novel Optoacoustic Device to Precisely Localize Endotracheal Tube Positioning in a Cadaver Model

Descriptive Note:

Technical Report,01 Sep 2014,31 Dec 2015

Corporate Author:

USAFSAM/ETS Wright-Patterson AFB

Report Date:

2016-11-02

Pagination or Media Count:

21.0

Abstract:

Due to the high likelihood of injury caused by improperly placed intubations and the movement of endotracheal tubes ETTs during transport, there is a great need for compact, user-friendly devices that are able to non-invasively detect the location of an ETT. Therefore, the objective of this study was to validate a novel optoacoustic device OAD as a feasible tool to precisely localize ETT position within the trachea using a cadaver model. Cadavers n24 were intubated with ETTs, and the location of the ETT was assessed using the novel OAD, the Prospiria trademark. A surgical dissection was then performed to ascertain the actual position of the internal optical input within the ETT. Results provide support that the novel OAD is able to accurately detect the location of the internal optical input within 2 cm of its actual location p0.001.

Subject Categories:

Distribution Statement:

APPROVED FOR PUBLIC RELEASE