Accession Number:
AD1013993
Title:
Evaluation of the Sensitivity and Signal Response of the DT-702 LiF:Mg,Cu,P TLD
Descriptive Note:
Technical Report
Corporate Author:
Uniformed Services University Of The Health Sciences Bethesda United States
Report Date:
2007-06-27
Pagination or Media Count:
90.0
Abstract:
The reliability and accuracy of a radiation dosimetry program is dependent on the intimate knowledge of the characteristics of the dosimetry material. In this study, the effects of pre-exposure sensitivity and post-exposure signal time on the Navys DT-702PD Harshaw Thermoluminescent Dosimeter 884041 are evaluated. The results of this study show that the DT-702 TLD has no statistically significant loss of sensitivity or loss of signal with up to 29 weeks of pre-exposure or post-exposure time. Therefore, the DT-702 could be in use for at least 29 weeks producing accurate results without any additional fade correction required.