Permeability Calibration of Cellophane Tubing Under Conditions of Simple Dialysis and Ultrafiltration.
Descriptive Note:
Progress rept.,
Corporate Author:
ARMY MEDICAL RESEARCH LAB FORT KNOX KY
Personal Author(s):
Report Date:
1974-05-07
Pagination or Media Count:
11.0
Abstract:
The percents escape of protein standards were shown to be directly proportional to their molecular weights. The 4,000 dalton standard demonstrated a loss of 40 after 24 hours of simple dialysis standards equal to or greater than 9,000 daltons demonstrated an almost negligible loss 2 or less. Dialysis membrane retained none of the 4,000 dalton standard during ultrafiltration and demonstrated escapes of 6 to 24 for standards with molecular weights between 9,000 and 45,000 daltons. Loss of the 67,000 and 150,000 dalton standards during ultrafiltration was negligible. Author