DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
AD0780565
Title:
Improvements to Automatic Oxygen Absorption Apparatus.
Descriptive Note:
Technical note,
Corporate Author:
EXPLOSIVES RESEARCH AND DEVELOPMENT ESTABLISHMENT WALTHAM ABBEY (ENGLAND)
Report Date:
1973-03-01
Pagination or Media Count:
10.0
Abstract:
Automatic apparatus for measurements of autoxidation at constant oxygen pressures normally relies on the make and break of mercury contacts. These are usually troublesome because of their sensitivity to small temperature changes and to contamination. These problems, as well as those connected with contacts in which a liquid meniscus can vary in shape, were successfully overcome with the aid of a stainless steel float and the elimination of the mercury itself, viz the electrical contact maker. Author
Distribution Statement:
APPROVED FOR PUBLIC RELEASE