Accession Number:
AD0769708
Title:
A Recoil Proton Telescope for Accurate Fast-Neutron Flux Measurement,
Descriptive Note:
Corporate Author:
BALLISTIC RESEARCH LABS ABERDEEN PROVING GROUND MD
Report Date:
1973-07-01
Pagination or Media Count:
41.0
Abstract:
A recoil proton telescope has been designed, built, and tested at the Radiation Laboratory, BRL. This device is capable of absolute fast-neutron flux measurement in the energy range 2-20 MeV to an accuracy of better than 3. Measurements of the telescope efficiencies at 2.60 and 14.30 MeV agree with several calculations to well within the estimated uncertainties. Thus, absolute flux measurements can be provided for the monitoring of irradiations and cross-section measurements. Author
Distribution Statement:
APPROVED FOR PUBLIC RELEASE