Accession Number:

AD0763333

Title:

Data Processing System for Standardization of Gamma and Neutron Calibrators,

Descriptive Note:

Corporate Author:

NAVAL ORDNANCE LAB WHITE OAK MD

Personal Author(s):

Report Date:

1973-04-10

Pagination or Media Count:

71.0

Abstract:

A data processing system was developed to process and analyze data taken in the field standardization of shore-based gamma and neutron calibrators. The gamma calibrator, the ANUDM-1A, is used to calibrate several types of gamma radiation detection instruments radiacs. The neutron calibrator, the MX-9335 XWO-1UDM, is used to calibrate a neutron detecting radiac set, the ANPDR-70. Two computer programs process data taken from sets of transfer standard instruments which are used to measure dose rate as a function of distance. One program is used to analyze the data, compute a linear regression line, and calculate several statistical parameters. In a second program, a series of tables are printed which list the distances at which specified dose rates are generated by the calibrator. The tables are dated and corrected for activity changes in the calibration sources. Author

Subject Categories:

  • Computer Programming and Software
  • Nuclear Instrumentation

Distribution Statement:

APPROVED FOR PUBLIC RELEASE