DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
AD0760536
Title:
Errors in Luminescence Decay Time Measurements Caused by Photomultiplier Saturation.
Descriptive Note:
Research rept. Jun-Sep 72,
Corporate Author:
AEROSPACE CORP EL SEGUNDO CALIF LAB OPERATIONS
Report Date:
1973-05-28
Pagination or Media Count:
16.0
Abstract:
A possible source of error in measuring luminescent decay times as a result of photomultiplier saturation is discussed. Photomultiplier saturation can occur by exceeding the maximum linear output current, which decreases with increased light pulse duration. For this reason it is necessary to calibrate the instrument response appropriate to the lifetime and light intensity being measured. Author
Distribution Statement:
APPROVED FOR PUBLIC RELEASE