DID YOU KNOW? DTIC has over 3.5 million final reports on DoD funded research, development, test, and evaluation activities available to our registered users. Click
HERE to register or log in.
Accession Number:
AD0672563
Title:
A CALORIMETER WITH INTERCHANGEABLE ABSORBERS FOR ABSORBED DOSE MEASUREMENTS,
Descriptive Note:
Corporate Author:
NAVAL RADIOLOGICAL DEFENSE LAB SAN FRANCISCO CALIF
Report Date:
1968-06-14
Pagination or Media Count:
15.0
Abstract:
A calorimeter for measuring absorbed dose is described that has interchangeable absorbers which can be placed within 2 cm of the source of radiation. The isothermal environment is maintained with boiling liquid nitrogen and a thermal switch permits rapid cooling of the detector. Absorbed dose measurements for four absorbers, C, Al, Si and Ge, are presented. Exposures at a 250 kV constant potential x-ray machine are compared to calculated values at four effective energies, 35, 65, 110 and 180 keV. Author
Distribution Statement:
APPROVED FOR PUBLIC RELEASE