ERROR ANALYSIS OF THE HUMIDITY-TEMPERATURE MEASURING SET, AN/TMQ-11.

Descriptive Note:

Instrumentation papers no. 140,

Corporate Author:

AIR FORCE CAMBRIDGE RESEARCH LABS L G HANSCOM FIELD MASS

Personal Author(s):

Report Date:

1968-03-01

Pagination or Media Count:

110.0

Abstract:

The report discusses the results of tests performed at AFCRL to determine how closely the ANTMQ-11 agrees with the ML-592. The analysis presented is based on 1769 observations obtained under field conditions from 1 November 1966 to 1 September 1967. The tests indicate that a negative bias error exists in the ANTMQ-11 lithium chloride Honeywell and Foxboro dewpoint sensors. If the dewcel readings are adjusted to compensate for this bias, dewpoint temperatures can be accepted as indicated approximately 90 percent of the time. Author