Abstract:

A method for measuring the dielectric properties of liquid specimens on a Q-meter at frequencies up to 100 MHz was developed by 1 assuming an equivalent circuit for a cylindrically shaped liquid-sample holder, 2 determining relevant circuit parameters by experiment, and 3 deriving the necessary equations taking full account of the effects of sample-holder inductance. Measurements made on two liquid samples are detailed and show that good accuracy is obtained when the standard liquid employed is a lossless dielectric with dielectric constant K near the K of the sample under test.

Subject Categories:

  • Test Facilities, Equipment and Methods
  • Electricity and Magnetism

Distribution Statement:

APPROVED FOR PUBLIC RELEASE