Title:
A COAXIAL CALORIMETER FOR USE AS A MICROWAVE POWER STANDARD,
Descriptive Note:
Corporate Author:
NATIONAL RESEARCH COUNCIL OF CANADA OTTAWA (ONTARIO) RADIO AND ELECTRICAL ENGINEERING DIV
Personal Author(s):
Report Date:
1964-09-28
Pagination or Media Count:
6.0
Abstract:
A coaxial calorimeter has been built for use in the comparison of power at high frequencies with dc resistance and voltage standards. The temperature rise of the metal jacket of a 50-ohm coaxial termination is measured with a thermopile when a known level of dc power is fed into the calorimeter. The high-frequency sensitivity of the calorimeter is obtained by modifying the dc sensitivity for 1 Input VSWR of the calorimeter, 2 Attenuation of a 6-inch length of thermally insulating input coaxial line, 3 The effect of current distribution in the load. The accuracy is estimated at better than 1 per cent below 4 Gcs with the largest errors due to the input VSWR of the calorimeter. Author