Accession Number:

AD0613979

Title:

CALIBRATION OF RADIOSONDES AT TEMPERATURES TO MINUS 100 DEGREES C,

Descriptive Note:

Corporate Author:

AMERICAN METEOROLOGICAL SOCIETY BOSTON MASS

Personal Author(s):

Report Date:

1964-11-01

Pagination or Media Count:

8.0

Abstract:

A method and apparatus to calibrate the thermistors of radiosondes are described. The method is based on cooling alcohol and toluene to -95 to -100C and a method for checking the readings of test thermometers at these temperatures against reference points. The method utilizes the familiar relationship between the sublimation point of dry ice and the external pressure.