Abstract:

The construction and calibration of a confocal wavemeter for the 94-GHz region is described. Material selection was predicted upon obtaining a frequency stability of 1 part in 100,000 per degree c over a 34 degree C temperature range. The system uses a 94-GHz signal generated from a crystal reference and phase-lock loop to calibrate the wavemeter. Several readout techniques are discussed 1 the conventional dip indication, 2 the pip obtained from the resonator used in the transmission mode, and 3 the measurement of the pip phase. The latter technique permits the resolution of 100 GHz or 1 part in 1,000,000. Author

Subject Categories:

  • Test Facilities, Equipment and Methods

Distribution Statement:

APPROVED FOR PUBLIC RELEASE