Title :   Drop Calibration of Accelerometers for Shock Measurement


Descriptive Note : Technical rept.


Corporate Author : DEFENSE SCIENCE AND TECHNOLOGY ORGANIZATION VICTORIA (AUSTRALIA) MARITIME PLATFORMS DIV


Personal Author(s) : Krelle, Andrew


Full Text : https://apps.dtic.mil/dtic/tr/fulltext/u2/a559387.pdf


Report Date : Aug 2011


Pagination or Media Count : 49


Abstract : This report details the Drop Calibration Method for determining the sensitivity of shock accelerometers. The Drop Calibration Method requires the accelerometer under calibration and either a traceable acceleration standard or a traceable force transducer standard to undergo a series of drop impacts. The sensitivity of the accelerometer is determined by comparing the outputs of the accelerometer and the standard. By appropriate characterisation of both the accelerometer's and the standard's data channels, prior to the calibration drops, the Drop Calibration Method ensures that the resultant sensitivity is traceable to National Association of Testing Authorities, Australia (NATA) standards.


Descriptors :   *ACCELEROMETERS , *CALIBRATION , *ERROR ANALYSIS , ACCELERATION , AUSTRALIA , DATA TRANSMISSION SYSTEMS , FORCE(MECHANICS) , IMPACT , MEASUREMENT , OUTPUT , SENSITIVITY , SHOCK , STANDARDS , TRACER STUDIES , TRANSDUCERS


Subject Categories : Test Facilities, Equipment and Methods