Accession Number : ADA255919


Title :   Test and Evaluation Report of the IVAC (Trademark) Volumetric Controller Model 280EE


Descriptive Note : Final rept.


Corporate Author : ARMY AEROMEDICAL RESEARCH LAB FORT RUCKER AL


Personal Author(s) : Bruckart, James E ; Licina, Joseph R ; Olding, Bill ; Quattlebaum, Martin


Full Text : https://apps.dtic.mil/dtic/tr/fulltext/u2/a255919.pdf


Report Date : Apr 1992


Pagination or Media Count : 60


Abstract : The IVAC Volumetric Controller, Model 280EE, was tested for electromagnetic interference/ compatibility in the UH-60A helicopter under the U.S. Army Program for Testing and Evaluation of Equipment for Aeromedical Operations. The tests were conducted using current military and industrial standards and procedures for electromagnetic interference/compatibility and human factors. The IVAC Model 280EE did not perform properly in the altitude chamber and excess electromagnetic emissions make it ineligible for an airworthiness release. Therefore, the unit is not considered compatible with the U.S. Army aeromedical aircraft.


Descriptors :   *TEST AND EVALUATION , *MODELS , *ELECTROMAGNETIC INTERFERENCE , EMISSION , COMPATIBILITY , OPERATION , STANDARDS , ALTITUDE , MEDICAL EQUIPMENT , ALTITUDE CHAMBERS , AIRWORTHINESS , ELECTROMAGNETIC COMPATIBILITY , CHAMBERS , VOLUMETRIC ANALYSIS , ARMY , RELEASE , INTERFERENCE , HELICOPTERS , HUMANS , AIRCRAFT


Subject Categories : Helicopters
      Radiofrequency Wave Propagation


Distribution Statement : APPROVED FOR PUBLIC RELEASE