Accession Number : AD0479767


Title :   STUDY OF MASS FLOWMETERS.


Descriptive Note : Final rept. 26 Jun 53-15 Jun 59,


Corporate Author : ARMOUR RESEARCH FOUNDATION CHICAGO IL


Personal Author(s) : Miesse, C. C.


Report Date : 15 JUL 1959


Pagination or Media Count : 133


Abstract : Part I contains a survey of mass flowmeters, including volumetric meters with automatic temperature-dependent density compensation, transverse momentum-type flowmeters, and other types of flowmeters which can be adapted to the measurement of mass flow. As a result of this survey it was concluded that volumetric meters with density compensation are inadequate for nonuniform mass flow, transverse momentum-type flowmeters are well-suited for mass flow measurement, and the more exotic device require further development before serious consideration can be given. Part II contains a detailed report of the calibration techniques and data obtained from the experimental investigation of three meters. (Author)


Descriptors :   *FLOWMETERS, RELIABILITY, INSTRUMENTATION, CALIBRATION, MEASUREMENT, ERRORS, DENSITY, GYROSCOPES, TRANSDUCERS, VENTURI TUBES, DETECTORS, CORIOLIS EFFECT, MOMENTUM, ACCELERATION, AUTOMATIC, MATHEMATICAL ANALYSIS, VELOCITY, SOUND GENERATORS, PRESSURE, FLUID FLOW.


Subject Categories : TEST FACILITIES, EQUIPMENT AND METEORS


Distribution Statement : APPROVED FOR PUBLIC RELEASE