Title :   A GAMMA-RAY DIGITAL DOSIMETRY SYSTEM,


Corporate Author : NAVAL RADIOLOGICAL DEFENSE LAB SAN FRANCISCO CA


Personal Author(s) : Taylor, R A


Report Date : 12 Jan 1965


Pagination or Media Count : 18


Abstract : A digital dosimeter system employing a scintillation detector and an analog to digital converter (ADC) was designed and constructed for the measurement of gamma-ray differential dose. The differential dose for gamma rays with energies up to 700 keV, and dose rates as low as 0.2 mr/hr, have been measured to better than plus or minus 11 percent with this instrument. The limitation on the accuracy of the electronics is of the order of plus or minus 1 percent. (Author)


Descriptors :   *GAMMA RAYS , DOSIMETERS , SCINTILLATION COUNTERS , ANALOG TO DIGITAL CONVERTERS , MEASUREMENT , PHOTOMULTIPLIER TUBES , DOSE RATE